Pepkio · Tool workspace

RCF ↔ RPM Rotor Converter

Convert RPM and ×g with rotor lookup, tri-radius output, protocol transfer, and copy-ready methods text.

Feedback

Help us improve this tool.

Need full analysis?

Selected: Beckman Coulter JA-20

Radius

RCF (×g)

rmin3,577.6 ×g
ravg7,826 ×g
rmax12,074.4 ×g

Force gradient (top → bottom)

rminravgrmax

Formula: RCF (×g) = 1.118×10⁻⁶ × r(mm) × RPM². Use rmax ×g for pelleting; rmin–rmax span matters for gradients. Protocol transfer matches RCF at rmax; target RPM is rounded to the nearest integer. With tube adapters, subtract adapter bottom thickness from rmax per manufacturer guidance.